- Inadequate Efficacy of Deep Brain Stimulation in a Patient with Parkinson's disease due to Partial Breakage of Electrode Lead
-
Chul Hyoung Lyoo
-
Department of Neurology, Youngdong Severance Hospital, Seoul;
Department of Neurosurgery, Yonsei University College of Medicinea, Seoul, Korea
- 파킨슨병 환자에서 전극선의 부분 손상으로 인한 심부뇌자극 효과 소실
-
류철형, 장진우a 이명식
-
영동세브란스병원 신경과, 연세대학교 의과대학 신경외과학교실a
- Abstract
- A patient with Parkinson's disease developed fluctuation in the deep brain stimulation (DBS) effect, an unpleasant left facial paresthesia and the left limb dystonia. Impedance of the right DBS was over 2000 ohm in three proximal contacts. Skull X-ray studies showed partial breakage of right electrode lead below the mastoid process. Partial electrode breakage must be considered when there is a deterioration of the DBS effect, an unexpected side effect of DBS, and an alteration of impedance.KeyWords:Deep brain stimulation, Parkinson disease